The Envair Bio 3+ is a totally enclosed 'glove box' that provides operator safety and product protection for the handling of biological agents up to and including ACDP Hazard Group 4 (subject to proper sitting, operation and maintenance) and has the following features:
Bio 3/1+ model for easy conversion between Class III mode (for biological agents up to and including Hazard Group 4) and Class 1 mode (for biological agents up to and including Hazard Group 3) where either application may be required.

| 3+ D | The Bio 3+D is a standard Class III microbiological safety cabinet which is designed to exhaust HEPA filtered air to atmosphere via a dedicated duct and remote extract fan. The workspace is totally enclosed and access to the work space is by the use of gloves or gauntlets set into the visor |
| 3/1+D | The Bio 3/1+D is a 'hybrid' cabinet that is designed to operate in Class III mode as above or in Class I mode. In Class III mode, access to the workspace is via gloves or gauntlets set into a removable closure panel. The conversion to Class I mode is easily accomplished by removing the closure panel and fixing blanking plate(s) over the inlet HEPA filter(s). Note: Commissioning and routine testing for the Bio 3/1+ must always be in both modes |
For more information please view
Product Sheet No. 103 1.12MB pdf.